Back Button U1252B Datasheet HTML 4Page - Keysight Technologies U1252B Datasheet HTML 5Page - Keysight Technologies U1252B Datasheet HTML 6Page - Keysight Technologies U1252B Datasheet HTML 7Page - Keysight Technologies U1252B Datasheet HTML 8Page - Keysight Technologies U1252B Datasheet HTML 9Page - Keysight Technologies U1252B Datasheet HTML 10Page - Keysight Technologies U1252B Datasheet HTML 11Page - Keysight Technologies U1252B Datasheet HTML 12Page - Keysight Technologies Next Button
Zoom Inzoom in Zoom Outzoom out
 8 / 15 page
background image
U1251B frequency sensitivity during voltage measurement
Frequency sensitivity and trigger level
Input range 3
Minimum sensitivity
(R.M.S. Sine Wave)
Trigger level for DC coupling
(Maximum input for specified
accuracy = 10 x Range or 1000 V)
20 Hz - 100 kHz
>100 kHz ~ 200 kHz
< 100 kHz
>100 kHz ~ 200 kHz
50.000 mV
10 mV
15 mV
10 mV
15 mV
500.00 mV
25 mV
35 mV
60 mV
70 mV
1000.0 mV
40 mV
50 mV
100 mV
150 mV
5.0000 V
0.25 V
0.5 V
0.5 V / 1.25 V (< 100 Hz)
0.6 V
50.000 V
2.5 V
5 V
5 V
6 V
500.00 V
25 V
N/A
50 V
N/A
1000.0 V
50 V
N/A
300 V
N/A
U1252B/U1253B frequency sensitivity during voltage measurement
Frequency sensitivity and trigger level
Input range 3
Minimum sensitivity
(RMS sine wave)
Trigger level for DC coupling
(Maximum input for specified
accuracy = 10 x range or 1000 V)
20 Hz ~ 200 kHz
>200 kHz ~ 500 kHz
< 100 kHz
>100 kHz ~ 500 kHz
50.000 mV
10 mV
25 mV
10 mV
25 mV
500.00 mV
70 mV
150 mV
70 mV
150 mV
1000.0 mV
120 mV
300 mV
120 mV
300 mV
5.0000 V
0.3 V
1.2 V
0.6 V
1.5 V
50.000 V
3 V
5 V
6 V
15 V
500.00 V
30 V < 100 kHz
N/A
60 V
N/A
1000.0 V
50 V < 100 kHz
N/A
120 V
N/A
Frequency sensitivity during current measurement
Input range
Minimum sensitivity (RMS sine wave)
20 Hz ~ 20 Khz
500.00 µA
100 µA
5000.0 µA
250 µA
50.000 mA
10 mA
440.00 mA
25 mA
5.0000 A
1 A
10.000 A
2.5 A
Frequency specifications 5
Range
Resolution
Accuracy
Min. input frequency 1
U1251B/2B
U1253B
99.999 Hz
0.001 Hz
0.02%+3 9
0.02%+3 9
1 Hz
999.99 Hz
0.01 Hz
0.02%+3, <600 kHz
0.02%+3, <600 kHz
9.9999 kHz
0.0001 kHz
99.999 kHz
0.001 kHz
999.99 kHz
0.01 kHz
1. The input signal is lower than the product
of 20000000 V x Hz (product voltage and
frequency); overload protection: 1000 V.
2. The multimeter will automatically select the
most appropriate range when making frequency
measurements.
3. Maximum input for specified accuracy = 10 x
range or 1000 V.
8 | Keysight | U1250 Series Handheld Digital Multimeter - Data Sheet


Ähnliche Teilenummer - U1252B

HerstellerTeilenummerDatenblattBauteilbeschribung
logo
Keysight Technologies
U1252B KEYSIGHT-U1252B Datasheet
7Mb / 42P
   For Electronic, Electrical and Industrial Process Testing
More results

Ähnliche Beschreibung - U1252B

HerstellerTeilenummerDatenblattBauteilbeschribung
logo
Agilent(Hewlett-Packard...
U1250 HP-U1250 Datasheet
2Mb / 19P
   Agilent U1250 Series Handheld Digital Multimeters
U1253B HP-U1253B Datasheet
562Kb / 14P
   Agilent U1250 Series Handheld Digital Multimeters
logo
Keysight Technologies
U1240 KEYSIGHT-U1240 Datasheet
1Mb / 9P
   Handheld Digital Multimeters
logo
B&K Precision Corporati...
388B BK-388B Datasheet
1Mb / 2P
   Handheld Digital Multimeters
logo
Keysight Technologies
U1281A KEYSIGHT-U1281A Datasheet
4Mb / 18P
   U1280 Series Handheld Digital Multimeters
U1282A KEYSIGHT-U1282A Datasheet
4Mb / 18P
   U1280 Series Handheld Digital Multimeters
logo
List of Unclassifed Man...
U1271A ETC2-U1271A Datasheet
3Mb / 22P
   U1270 Series Handheld Digital Multimeters
logo
Keysight Technologies
U1271A KEYSIGHT-U1271A Datasheet
4Mb / 22P
   U1270 Series Handheld Digital Multimeters
U1273AX KEYSIGHT-U1273AX Datasheet
4Mb / 22P
   U1270 Series Handheld Digital Multimeters
U1272A KEYSIGHT-U1272A Datasheet
4Mb / 22P
   U1270 Series Handheld Digital Multimeters
More results


Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15


Datenblatt Download

Go To PDF Page


Link URL




Privatsphäre und Datenschutz
ALLDATASHEETDE.COM
War ALLDATASHEET hilfreich?  [ DONATE ] 

Über Alldatasheet   |   Werbung   |   Kontakt   |   Privatsphäre und Datenschutz   |   Linktausch   |   Hersteller
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com